MEASURING OXYGEN
Dynamic
Fault-proof
Absolute
Pollutants
COMPOST FORMATION
MEASURING CARBON DIOXIDE
AIR QUALITY

   
MEASURING OXYGEN:
Pollutants
 

 

Pollutants will destroy the oxygen sensor

Since the oxygen sensor contains zirconium dioxide and platinum, the following substances can destroy it:
  • Heavy metals
  • Sulphuric compounds
  • Silicone vapors
  • Fluorine
  • NH3 (as of 1000 ppm)
  • Halogenated hydrocarbons (as of 100 ppm)
  • Phosphate ester
  • Chlorine
  • SF6
  • Carbons
  • Salts
  • Long time in reducing atmosphere

Dust, vibrations, dirt, humidity, oils, greases, furnace cleaning agents, heavy gas oil, pyrolysis gases and silicon oxide (generated when silicones burn in a furnace) decrease the service life of the oxygen sensor.

No responsibility is taken for the completeness of this list.

 
 